A1c, GMI and average glucose: why they don't always agree
Your lab A1c, your CGM's GMI and your app's estimated A1C look like versions of one number. Each is built differently, and knowing how helps you read a mismatch calmly and usefully.
To convert blood glucose from mmol/L to mg/dL, multiply by 18 (more precisely 18.016), and divide by 18 to go back. Here's where that number comes from, where common thresholds come from, and how A1c units compare.
Your meter reads 7.8. A friend in Seattle says she tries to keep her glucose between 70 and 180. You’re both describing blood glucose, in two different units. Canada uses millimoles per litre (mmol/L) and the United States uses milligrams per decilitre (mg/dL). One number links them. Multiply mmol/L by 18 (18.016, to be exact) to get mg/dL, or divide mg/dL by the same number to get mmol/L. Your 7.8 mmol/L is about 140 mg/dL, and her range works out to 3.9-10.0 mmol/L (70-180 mg/dL).
The arithmetic is simple. The confusion comes from rounding, from cut-offs translated from one unit into the other, from guidelines that set different targets in different countries, and from A1c, which has two units of its own. This article covers each, with the formulas written out so you can check any number yourself.
mg/dL = mmol/L × 18.016, and mmol/L = mg/dL ÷ 18.016, rounded to one decimal in mmol/L and to a whole number in mg/dL.
Both units describe concentration, meaning how much glucose is dissolved in a set volume of blood. They differ in how the glucose is counted.
One unit weighs the glucose and the other counts it. A roll of identical coins can be described by its weight or by how many coins it holds, and if you know what one coin weighs, either number gives you the other. For glucose, that weight is the molar mass, the mass of one mole of molecules.
A glucose molecule is made of 6 carbon, 12 hydrogen and 6 oxygen atoms (C6H12O6). Its molar mass is 180.16 grams per mole, and the US National Institute of Standards and Technology lists it as 180.1559 1. The conversion then takes three steps.
That makes 1 mmol/L equal to 18.016 mg/dL, which gives two formulas.
For example, 7.0 mmol/L × 18.016 = 126.1, so 7.0 mmol/L is 126 mg/dL. In the other direction, 100 mg/dL ÷ 18.016 = 5.55, which rounds to 5.6 mmol/L.
You’ll often see the factor rounded to 18. That changes the result by less than 0.1%, which moves the rounded answer by 1 mg/dL at most. For example, 7.8 mmol/L is 140.4 mg/dL with 18 and 140.5 with 18.016. The calculations in this article use 18.016.
To convert in your head, multiply mmol/L by 2 and then by 9 (7 × 2 = 14, and 14 × 9 = 126). For the reverse, halve the mg/dL value and divide by 9 (180 ÷ 2 = 90, and 90 ÷ 9 = 10). The factor works only for glucose. Cholesterol and other substances have molecules of different weights, so each has its own factor.
In plain terms: Multiply mmol/L by 18 to get mg/dL. Divide mg/dL by 18 to get mmol/L. Using the exact factor, 18.016, moves the rounded answer by 1 mg/dL at most.
Canada works in mmol/L. Diabetes Canada’s clinical practice guidelines give glucose values in mmol/L. To reach an A1c of 7.0% or less, they suggest aiming for 4.0-7.0 mmol/L when fasting or before meals and 5.0-10.0 mmol/L two hours after eating 2. In 2010, Health Canada told manufacturers that a blood glucose meter applying for a Canadian licence had to report results only in mmol/L 3.
The United States works in mg/dL. Guidance from the US Food and Drug Administration (FDA) on home blood glucose meters says meters intended for use in the US should report results in mg/dL 4. The American Diabetes Association (ADA) Standards of Care put mg/dL first, with mmol/L in parentheses. In the 2024 edition, the before-meal goal for many nonpregnant adults is 4.4-7.2 mmol/L (80-130 mg/dL) 5.
Those two before-meal goals differ in more than their units. Diabetes Canada’s 4.0-7.0 mmol/L works out to about 72-126 mg/dL, while the ADA’s range starts and ends higher 2, 5. When numbers from two countries don’t line up, convert first, then check whether the targets themselves differ.
International consensus documents often print both units. The 2019 international consensus on time in range writes mg/dL first, with mmol/L in parentheses 6. A 2017 joint position statement of the ADA and the European Association for the Study of Diabetes (EASD) put mmol/L first, starting with its title, which calls for glucose values below 3.0 mmol/L (54 mg/dL) to be reported in clinical trials 7.
The cut-offs used for time in range were agreed in consensus reports published in 2017. In one, a steering committee representing organizations that included the ADA, the Endocrine Society and JDRF defined outcome measures for type 1 diabetes beyond A1c 8.
The joint ADA and EASD statement from the same year described a glucose of 3.9 mmol/L (70 mg/dL) or less as an alert value, and a level below 3.0 mmol/L (54 mg/dL) as low enough to signal serious, clinically important hypoglycemia 7. The 2019 time-in-range consensus built its targets on the same cut-offs 6, and the ADA’s classification of hypoglycemia reprints the 2017 definitions 5, 8. For pregnancy, the time-in-range consensus uses a narrower range of 3.5-7.8 mmol/L (63-140 mg/dL) 6.
Values such as 70, 180 and 250 mg/dL are round numbers. Divided by 18.016, 70 and 250 become 3.885 and 13.877 mmol/L, which round to 3.9 and 13.9. Two pairs line up neatly in both units, because 3 × 18 = 54 and 10 × 18 = 180.
The table lists values you’ll meet often. The third column shows the exact product of the mmol/L value and 18.016.
| mmol/L | mg/dL | × 18.016 | Where you’ll see it |
|---|---|---|---|
| 3.0 | 54 | 54.0 | Level 2 low |
| 3.9 | 70 | 70.3 | Level 1 low; bottom of range |
| 5.3 | 95 | 95.5 | Pregnancy fasting goal |
| 6.7 | 120 | 120.7 | Pregnancy 2-hour goal |
| 7.0 | 126 | 126.1 | Diagnosis, fasting |
| 7.8 | 140 | 140.5 | Top of pregnancy range |
| 10.0 | 180 | 180.2 | Top of target range |
| 11.1 | 200 | 200.0 | Diagnosis, 2-hour or random |
| 13.9 | 250 | 250.4 | Level 2 high |
The levels and ranges come from the consensus documents above 6–8. The pregnancy goals are Diabetes Canada’s for people whose diabetes began before pregnancy 9. The diagnostic cut-offs, also Diabetes Canada’s, apply to venous blood measured in a laboratory, and without clear symptoms a result in the diabetes range needs a repeat lab test on another day 10.
For 6.7 and 7.8 mmol/L, the exact products round to 121 and 141 mg/dL. The usual pairs work from the other direction. Starting from 120 and 140 mg/dL gives 6.66 and 7.77 mmol/L, which round to 6.7 and 7.8.
Glucose in mmol/L is normally written to one decimal place and mg/dL as a whole number, which is how the consensus documents print them 6, 7. Converting 70 mg/dL gives 3.885 mmol/L, and rounding to one decimal gives 3.9. Writing 3.89 would suggest a precision that home meters don’t have. FDA guidance asks that 95% of a meter’s results fall within 15% of the results from a comparison method across its measuring range 4. At 3.9 mmol/L, 15% is about 0.6 mmol/L (10.5 mg/dL), far larger than any rounding.
Rounding explains a few small mismatches you may notice.
These differences are smaller than the measurement error that meter standards allow 4.
A1c, the share of hemoglobin with glucose attached, is reported in two units that come from two measurement systems.
Canadian guidelines and US laboratories use percent 2, 11. Through the National Glycohemoglobin Standardization Program (NGSP), the percent scale is tied to A1c results from two landmark trials, the Diabetes Control and Complications Trial (DCCT) and the UK Prospective Diabetes Study (UKPDS). Care guidelines from the ADA and other diabetes organizations relate directly to these numbers 11.
The mmol/mol unit is based on a reference method of the International Federation of Clinical Chemistry and Laboratory Medicine (IFCC), published in 2002. It defines A1c as hemoglobin with glucose stably attached at one exact spot, the start of its beta chain. The method cuts hemoglobin into fragments with an enzyme and measures the ratio of glycated to non-glycated fragments from that spot. Because it is more specific, it gives lower results than most routine methods in use at the time 12. In 2007, the IFCC recommended reporting the result as millimoles of A1c per mole of hemoglobin, or mmol/mol 11.
Between 2001 and 2003, the IFCC reference laboratories and the US, Japanese and Swedish standardization networks measured the same pooled blood samples in four studies. The relationship between IFCC and NGSP results was a straight line that didn’t change significantly from one study to the next, published as NGSP (%) = 0.915 × IFCC (%) + 2.15 13. That paper gave IFCC results as a percentage, and multiplying by 10 turns them into mmol/mol 11, 13. Rearranged, the equation gives two conversions.
The NGSP states the master equation with one more digit, NGSP = 0.09148 × IFCC + 2.152 with IFCC in mmol/mol, and writes the reverse as IFCC = 10.93 × NGSP − 23.50 11. For the values in the table below, every version gives the same whole number.
Two things follow. A1c in mmol/mol isn’t the percent times 10, since 7.0% is 53 mmol/mol. And each percentage point is worth about 11 mmol/mol, so a fall from 8.0% to 7.0% is a fall from 64 to 53 mmol/mol.
| A1c (%) | A1c (mmol/mol) |
|---|---|
| 5.7 | 39 |
| 6.0 | 42 |
| 6.5 | 48 |
| 7.0 | 53 |
| 8.0 | 64 |
| 9.0 | 75 |
Values are calculated with IFCC = (NGSP − 2.15) × 10.929 and rounded to whole numbers.
In 2010, a consensus committee of the ADA, the EASD, the International Diabetes Federation, the IFCC and the International Society for Pediatric and Adolescent Diabetes recommended that laboratories worldwide report A1c in both units, giving mmol/mol with no decimals and percent with one decimal 14. Practice varies. Diabetes Canada uses percent, with 6.5% (48 mmol/mol) or more as one of its criteria for diabetes, 6.0-6.4% (42-46 mmol/mol) as one marker of prediabetes 10 and 7.0% (53 mmol/mol) or less as the target for most people with type 1 or type 2 diabetes 2. The ADA writes percent with mmol/mol in parentheses 5, and some countries report mmol/mol alone or alongside percent 11.
In North America, the unit a meter shows depends on where it’s sold. FDA guidance says meters for use in the US should report results in mg/dL, as plasma-equivalent values 4. Health Canada’s 2010 notice went further. A meter applying for a Canadian licence had to report only in mmol/L, with a unit that neither the user nor a drop or battery change could alter, and any software supplied with it had to show mmol/L only 3.
Those rules followed real problems. Health Canada noted that the unit shown on some meters could change by accident when the device was dropped, when batteries were changed or with no apparent cause 3. In 2005, Abbott recalled several meter models in the US because users could switch the unit from mg/dL to mmol/L without meaning to. According to the FDA’s recall record, results could then be misread, which could lead to undertreatment and high glucose. The company’s fix included locking the unit on newly made meters 15.
Phone and CGM apps display a unit as well, so the same check applies to them. And because US and Canadian meters are set up for different units, a meter bought in another country may show a different one.
Because the two scales differ by a factor of about 18, a number read in the wrong unit is far off. A reading of 100 is ordinary in mg/dL (5.6 mmol/L) and would be extremely high in mmol/L. A reading of 5.6 is ordinary in mmol/L and would be dangerously low in mg/dL. Health Canada warned that a person who overestimates their glucose might take too much insulin and have a severe low, with loss of consciousness or a seizure 3. The recall above shows the reverse risk, undertreatment and high glucose 15.
Published data on how often unit mix-ups happen, or how often they cause harm, are scarce. The evidence above comes from regulators’ notices and a recall, which describe the problem without counting cases. The practical lessons are simple.
Multiply mmol/L by 18 (18.016, to be exact) to get mg/dL, or divide mg/dL by the same number to get mmol/L. For example, 7.0 mmol/L is 126 mg/dL. Rounding the factor to 18 moves the rounded answer by 1 mg/dL at most.
Both describe glucose concentration, but mmol/L counts the molecules, in millimoles per litre, while mg/dL weighs them, in milligrams per decilitre. Glucose's molar mass is 180.16 grams per mole 1. So one millimole weighs 180.16 mg, and since a litre holds 10 decilitres, 1 mmol/L equals 18.016 mg/dL.
Canada uses mmol/L, and the United States uses mg/dL. Diabetes Canada's guidelines give glucose values in mmol/L, while FDA guidance says meters intended for use in the US should report results in mg/dL 4. A meter bought in another country may show a different unit.
Subtract 2.15 from the percent and multiply by 10.929. So 6.5% is 48 mmol/mol, 7.0% is 53, and each percentage point is worth about 11 mmol/mol. A1c in mmol/mol isn't the percent times 10.
Some meters once did. Health Canada noted that units could change by accident when a meter was dropped or batteries were changed, and in 2010 told makers of meters seeking a licence to report only in mmol/L 3. Check the unit label when a reading looks far off.
Glyve supports both mmol/L and mg/dL. Choose the unit your meter and care team use, so the numbers on your screen match the ones in your clinic notes. Readings can come from Apple Health, Dexcom Share or LibreLinkUp, or you can log them by hand, by voice or with Siri. When you enter a reading yourself, check that it’s in the unit Glyve is showing. A value that looks about 18 times too high or too low points to a unit mismatch somewhere. Glyve shows time in range, estimated A1C (GMI) and averages for the period you view. Before comparing them with a report from another source, check that both use the same unit and the same target range, for example 3.9-10.0 mmol/L (70-180 mg/dL).